nday
CVE-2025-14096

Credential Disclosure vulnerability in Radiometer Products

Radiometer Medical Aps / ABL90 FLEX and ABL90 FLEX PLUS Analyzers、AQT90 FLEX Analyzers、ABL800 BASIC and ABL800 FLEX analyzers

揭露時間:。nday 尚未取得修補依據,不能據此認為沒有修補。四個 PoC 來源皆未收錄。目前未列入 CISA KEV。

why these tags

nday derives these tags by fixed rules from the facts below; NVD, the CNA and other sources do not assign them. Vocabulary and rules: ~/tags.

improper privilege management
CWE-250, stated by the CNA Radiometer 與 NVD, republishing 46b595e9-1acc-41cb-9398-adaf98d37a9b
hard-coded credentials
CWE-798, stated by the CNA Radiometer 與 NVD, republishing 46b595e9-1acc-41cb-9398-adaf98d37a9b
local
AV:L in CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Radiometer
no authentication required
PR:N in CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Radiometer
no user interaction
UI:N in CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Radiometer
目前階段
DIS
揭露時間
CVSS
8.4 HIGH v3.1

生命週期

橫軸 UTC;點=精確到分鐘,寬條=僅日期,空心=不明或推斷。
時間不明
DIS揭露
2025-12-17T12:19Z
PAT修補
尚無紀錄
ANA分析
尚無紀錄
POCPoC
尚無紀錄
ITW遭利用
尚無紀錄
KEVKEV
尚無紀錄

摘要

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.

Other related CVE's are CVE-2025-14095 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure:

Attacker requires physical access to the analyzer.

Temporary work Around: Only authorized people can physically access the analyzer.

Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status:

Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Original text of the CVE record (cvelist), shown as published.

Exploit conditions

What an attack needs, read from the CVSS vector.
Attack vector AV:L
Local: needs a local or SSH session, or a user tricked into acting on the machine
Attack complexity AC:L
Low: no special conditions; the attacker can expect repeatable success
Privileges required PR:N
None: no account or authorization needed
User interaction UI:N
None: no action by anyone other than the attacker

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

CVSS 3.1 vector from the CVE record, scored by the CNA Radiometer.

事件紀錄

+DIS揭露

CVE 紀錄由 Radiometer 發布。CVSS 8.4,屬高風險等級。

來源 cvelist · nday 於 批次匯入 · 原始
{
  "id": 361502,
  "type": "disclosed",
  "source": "cvelist",
  "url": "https://www.cve.org/CVERecord?id=CVE-2025-14096",
  "occurred_at": "2025-12-17T12:19:27.489Z",
  "observed_at": "2026-11-13T05:38:57.035223Z",
  "is_backfill": true,
  "dedupe_key": "",
  "detail": {
    "cvss": {
      "source": "cna",
      "vector": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
      "version": "3.1",
      "base_score": 8.4,
      "base_severity": "HIGH"
    },
    "ssvc": {
      "timestamp": "2025-12-17T14:42:57.843796Z",
      "automatable": "no",
      "exploitation": "none",
      "technical_impact": "total"
    },
    "state": "PUBLISHED",
    "date_published": "2025-12-17T12:19:27.489Z",
    "assigner_short_name": "Radiometer"
  }
}

Report an error

report an error or omission

Something on this page wrong or missing? Tell the editors; if they accept it, this page's revision log says so publicly. Upstream data is not changed.

研究員討論

近 7 天有 0 位研究員提及

名單內研究員在 X 上提及此 CVE 的貼文,照原文呈現,nday 不查核。

尚未收錄名單內研究員在 X 上提及此 CVE 的貼文。

Discussion

0 comments

No comments yet.