nday
CVE-2025-14096

Credential Disclosure vulnerability in Radiometer Products

Radiometer Medical Aps / ABL90 FLEX and ABL90 FLEX PLUS Analyzers, AQT90 FLEX Analyzers, ABL800 BASIC and ABL800 FLEX analyzers

CVE 공개: . nday가 아직 패치 근거를 수집하지 않았습니다. 패치가 없다는 뜻은 아닙니다. PoC 출처 4곳 모두 미등재. CISA KEV 미등재.

why these tags

nday derives these tags by fixed rules from the facts below; NVD, the CNA and other sources do not assign them. Vocabulary and rules: ~/tags.

improper privilege management
CWE-250, stated by the CNA Radiometer 및 NVD, republishing 46b595e9-1acc-41cb-9398-adaf98d37a9b
hard-coded credentials
CWE-798, stated by the CNA Radiometer 및 NVD, republishing 46b595e9-1acc-41cb-9398-adaf98d37a9b
local
AV:L in CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Radiometer
no authentication required
PR:N in CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Radiometer
no user interaction
UI:N in CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Radiometer
현재 단계
DIS
공개
CVSS
8.4 HIGH v3.1

생명주기

가로축 UTC. 점 = 분 단위, 넓은 막대 = 날짜만, 빈 표시 = 불명 또는 추정.
시점 알 수 없음
DIS공개
2025-12-17T12:19Z
PAT패치
아직 기록 없음
ANA분석
아직 기록 없음
POCPoC
아직 기록 없음
ITW악용
아직 기록 없음
KEVKEV
아직 기록 없음

요약

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.

Other related CVE's are CVE-2025-14095 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure:

Attacker requires physical access to the analyzer.

Temporary work Around: Only authorized people can physically access the analyzer.

Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status:

Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Original text of the CVE record (cvelist), shown as published.

Exploit conditions

What an attack needs, read from the CVSS vector.
Attack vector AV:L
Local: needs a local or SSH session, or a user tricked into acting on the machine
Attack complexity AC:L
Low: no special conditions; the attacker can expect repeatable success
Privileges required PR:N
None: no account or authorization needed
User interaction UI:N
None: no action by anyone other than the attacker

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

CVSS 3.1 vector from the CVE record, scored by the CNA Radiometer.

이벤트 로그

+DISCVE 공개

CVE 레코드 공개(CNA: Radiometer). CVSS 8.4(높음).

출처 cvelist · nday 일괄 수집 · 원본
{
  "id": 361502,
  "type": "disclosed",
  "source": "cvelist",
  "url": "https://www.cve.org/CVERecord?id=CVE-2025-14096",
  "occurred_at": "2025-12-17T12:19:27.489Z",
  "observed_at": "2026-11-13T05:38:57.035223Z",
  "is_backfill": true,
  "dedupe_key": "",
  "detail": {
    "cvss": {
      "source": "cna",
      "vector": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
      "version": "3.1",
      "base_score": 8.4,
      "base_severity": "HIGH"
    },
    "ssvc": {
      "timestamp": "2025-12-17T14:42:57.843796Z",
      "automatable": "no",
      "exploitation": "none",
      "technical_impact": "total"
    },
    "state": "PUBLISHED",
    "date_published": "2025-12-17T12:19:27.489Z",
    "assigner_short_name": "Radiometer"
  }
}

Report an error

report an error or omission

Something on this page wrong or missing? Tell the editors; if they accept it, this page's revision log says so publicly. Upstream data is not changed.

연구자 논의

최근 7일간 연구자 0명이 언급

목록의 연구자가 X에서 이 CVE를 언급한 게시물입니다. 원문 그대로 보여 주며 nday는 검증하지 않습니다.

명단의 연구자가 X에서 이 CVE를 언급한 게시물은 아직 수집되지 않았습니다.

Discussion

0 comments

No comments yet.