nday
CVE-2025-4394

Medtronic MyCareLink Patient Monitor Unencrypted Filesystem Vulnerability

Medtronic / MyCareLink Patient Monitor 24950, MyCareLink Patient Monitor 24952

Disclosed . nday has no fix record for it yet; that does not mean no patch exists. None of the four PoC sources lists it. Not in CISA KEV.

why these tags

nday derives these tags by fixed rules from the facts below; NVD, the CNA and other sources do not assign them. Vocabulary and rules: ~/tags.

cryptographic failure
CWE-312, stated by the CNA Medtronic and NVD, republishing [email protected]
physical
AV:P in CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Medtronic
no authentication required
PR:N in CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Medtronic
no user interaction
UI:N in CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.1) from the CNA Medtronic
stage
DIS
disclosed
cvss
6.8 MEDIUM v3.1

Lifecycle

x axis UTC; dot = exact minute, wide bar = date only, hollow = unknown or inferred.
time unknown
DISdisclosed
2025-07-24T03:26Z
PATpatch
no record yet
ANAanalysis
no record yet
POCPoC
no record yet
ITWexploited
no record yet
KEVKEV
no record yet

Summary

Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files.

This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025

Original text of the CVE record (cvelist), shown as published.

Exploit conditions

What an attack needs, read from the CVSS vector.
Attack vector AV:P
Physical: needs physical access to the device
Attack complexity AC:L
Low: no special conditions; the attacker can expect repeatable success
Privileges required PR:N
None: no account or authorization needed
User interaction UI:N
None: no action by anyone other than the attacker

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

CVSS 3.1 vector from the CVE record, scored by the CNA Medtronic.

Event log

+DISdisclosed

CVE record published by Medtronic. CVSS 6.8, medium.

src cvelist · nday , bulk import · raw
{
  "id": 352613,
  "type": "disclosed",
  "source": "cvelist",
  "url": "https://www.cve.org/CVERecord?id=CVE-2025-4394",
  "occurred_at": "2025-07-24T03:26:06.706Z",
  "observed_at": "2026-11-13T05:38:48.939217Z",
  "is_backfill": true,
  "dedupe_key": "",
  "detail": {
    "cvss": {
      "source": "cna",
      "vector": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
      "version": "3.1",
      "base_score": 6.8,
      "base_severity": "MEDIUM"
    },
    "ssvc": {
      "timestamp": "2025-07-24T13:19:43.967176Z",
      "automatable": "no",
      "exploitation": "none",
      "technical_impact": "total"
    },
    "state": "PUBLISHED",
    "date_published": "2025-07-24T03:26:06.706Z",
    "assigner_short_name": "Medtronic"
  }
}

Report an error

report an error or omission

Something on this page wrong or missing? Tell the editors; if they accept it, this page's revision log says so publicly. Upstream data is not changed.

Researcher discussion

0 researchers mentioned this in the last 7 days

Posts on X by researchers on nday's list that mention this CVE, shown as posted. nday does not check them.

nday has not collected any post on X by listed researchers that mentions this CVE.

Discussion

0 comments

No comments yet.