nday
CVE-2022-43557

BD BodyGuard™ Pumps – RS-232 Interface Vulnerability

Becton, Dickinson and Company (BD) / BodyGuard™ Pump

Disclosed . nday has no fix record for it yet; that does not mean no patch exists. None of the four PoC sources lists it. Not in CISA KEV.

why these tags

nday derives these tags by fixed rules from the facts below; NVD, the CNA and other sources do not assign them. Vocabulary and rules: ~/tags.

authentication bypass
CWE-287, stated by NVD's analysts
physical
AV:P in CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H (CVSS 3.1) from the CNA BD
no authentication required
PR:N in CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H (CVSS 3.1) from the CNA BD
no user interaction
UI:N in CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H (CVSS 3.1) from the CNA BD
stage
DIS
disclosed
cvss
5.3 MEDIUM v3.1

Lifecycle

x axis UTC; dot = exact minute, wide bar = date only, hollow = unknown or inferred.
time unknown
DISdisclosed
2022-12-05T00:00Z
PATpatch
no record yet
ANAanalysis
no record yet
POCPoC
no record yet
ITWexploited
no record yet
KEVKEV
no record yet

Summary

The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.

Original text of the CVE record (cvelist), shown as published.

Exploit conditions

What an attack needs, read from the CVSS vector.
Attack vector AV:P
Physical: needs physical access to the device
Attack complexity AC:H
High: success depends on conditions outside the attacker's control, such as gathering details of the target or winning a race
Privileges required PR:N
None: no account or authorization needed
User interaction UI:N
None: no action by anyone other than the attacker

CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H

CVSS 3.1 vector from the CVE record, scored by the CNA BD.

Event log

+DISdisclosed

CVE record published by BD. CVSS 5.3, medium.

src cvelist · nday , bulk import · raw
{
  "id": 275146,
  "type": "disclosed",
  "source": "cvelist",
  "url": "https://www.cve.org/CVERecord?id=CVE-2022-43557",
  "occurred_at": "2022-12-05T00:00:00Z",
  "observed_at": "2026-11-13T05:37:47.986087Z",
  "is_backfill": true,
  "dedupe_key": "",
  "detail": {
    "cvss": {
      "source": "cna",
      "vector": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H",
      "version": "3.1",
      "base_score": 5.3,
      "base_severity": "MEDIUM"
    },
    "ssvc": {
      "timestamp": "2025-04-23T15:46:47.466198Z",
      "automatable": "no",
      "exploitation": "none",
      "technical_impact": "partial"
    },
    "state": "PUBLISHED",
    "date_published": "2022-12-05T00:00:00.000Z",
    "assigner_short_name": "BD"
  }
}

Report an error

report an error or omission

Something on this page wrong or missing? Tell the editors; if they accept it, this page's revision log says so publicly. Upstream data is not changed.

Researcher discussion

0 researchers mentioned this in the last 7 days

Posts on X by researchers on nday's list that mention this CVE, shown as posted. nday does not check them.

nday has not collected any post on X by listed researchers that mentions this CVE.

Discussion

0 comments

No comments yet.