Siemens RAPIDLab 1200 systems RAPIDPoint 400 systems…: hard-coded credentialsTitle composed by nday; the CVE record has none.

Siemens AG / RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems

Disclosed . nday has no fix record for it yet; that does not mean no patch exists. None of the four PoC sources lists it. Not in CISA KEV.

why these tags

nday derives these tags by fixed rules from the facts below; NVD, the CNA and other sources do not assign them. Vocabulary and rules: ~/tags.

hard-coded credentials
CWE-798, stated by the CNA siemens, NVD's analysts, and NVD, republishing [email protected]
remote, unauthenticated
AV:N, PR:N, UI:N in CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.0) from NVD's analysts
network
AV:N in CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.0) from NVD's analysts
no authentication required
PR:N in CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.0) from NVD's analysts
no user interaction
UI:N in CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (CVSS 3.0) from NVD's analysts
stage
DIS
disclosed
cvss
no score yet

Lifecycle

x axis UTC; dot = exact minute, wide bar = date only, hollow = unknown or inferred.
time unknown
DISdisclosed
2018-06-26T18:00Z
PATpatch
no record yet
ANAanalysis
no record yet
POCPoC
no record yet
ITWexploited
no record yet
KEVKEV
no record yet

Summary

A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

Original text of the CVE record (cvelist), shown as published.

Event log

+DISdisclosed

CVE record published by siemens.

src cvelist · nday , bulk import · raw
{
  "id": 181512,
  "type": "disclosed",
  "source": "cvelist",
  "url": "https://www.cve.org/CVERecord?id=CVE-2018-4846",
  "occurred_at": "2018-06-26T18:00:00Z",
  "observed_at": "2026-11-13T05:36:42.797974Z",
  "is_backfill": true,
  "dedupe_key": "",
  "detail": {
    "state": "PUBLISHED",
    "date_published": "2018-06-26T18:00:00.000Z",
    "assigner_short_name": "siemens"
  }
}

Report an error

report an error or omission

Something on this page wrong or missing? Tell the editors; if they accept it, this page's revision log says so publicly. Upstream data is not changed.

Researcher discussion

0 researchers mentioned this in the last 7 days

Posts on X by researchers on nday's list that mention this CVE, shown as posted. nday does not check them.

nday has not collected any post on X by listed researchers that mentions this CVE.

Discussion

0 comments

No comments yet.