Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

Dräger / Infinity Acute Care System, Standalone Infinity M540 patient monitor

Fecha de divulgación: . nday aún no tiene constancia de una corrección; eso no significa que no exista parche. Ninguna de las cuatro fuentes de PoC la recoge. No figura en el KEV de CISA.

why these tags

nday derives these tags by fixed rules from the facts below; NVD, the CNA and other sources do not assign them. Vocabulary and rules: ~/tags.

remote, unauthenticated
AV:N, PR:N, UI:N in CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N (CVSS 4.0) from the CNA VulnCheck
network
AV:N in CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N (CVSS 4.0) from the CNA VulnCheck
no authentication required
PR:N in CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N (CVSS 4.0) from the CNA VulnCheck
no user interaction
UI:N in CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N (CVSS 4.0) from the CNA VulnCheck
etapa
DIS
divulgación
cvss
8.8 HIGH v4.0

Ciclo de vida

Eje x en UTC; punto = minuto exacto, barra ancha = solo fecha, hueco = desconocida o inferida.
fecha desconocida
DISdivulgada
06-02T14:00Z
PATparcheada
sin registro aún
ANAanalizada
sin registro aún
POCPoC
sin registro aún
ITWexplotada
sin registro aún
KEVen KEV
sin registro aún

Descripción

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Original text of the CVE record (cvelist), shown as published.

Exploit conditions

What an attack needs, read from the CVSS vector.
Attack vector AV:N
Network: the attacker only needs to reach it over the network
Attack complexity AC:L
Low: no target-specific security measures to get around
Attack requirements AT:N
None: success does not depend on how the target is deployed or running
Privileges required PR:N
None: no account or authorization needed
User interaction UI:N
None: no action by anyone other than the attacker

CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N

CVSS 4.0 vector from the CVE record, scored by the CNA VulnCheck.

Registro de eventos

+DISdivulgación

VulnCheck publicó el registro CVE. CVSS 8.8, severidad alta.

fuente cvelist · nday , importación masiva · en bruto
{
  "id": 210523,
  "type": "disclosed",
  "source": "cvelist",
  "url": "https://www.cve.org/CVERecord?id=CVE-2019-25719",
  "occurred_at": "2026-06-02T14:00:52.578Z",
  "observed_at": "2026-11-13T05:37:01.102615Z",
  "is_backfill": true,
  "dedupe_key": "",
  "detail": {
    "cvss": {
      "source": "cna",
      "vector": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N",
      "version": "4.0",
      "base_score": 8.8,
      "base_severity": "HIGH"
    },
    "ssvc": {
      "timestamp": "2026-06-02T14:56:18.300929Z",
      "automatable": "yes",
      "exploitation": "none",
      "technical_impact": "partial"
    },
    "state": "PUBLISHED",
    "date_published": "2026-06-02T14:00:52.578Z",
    "assigner_short_name": "VulnCheck"
  }
}

Report an error

report an error or omission

Something on this page wrong or missing? Tell the editors; if they accept it, this page's revision log says so publicly. Upstream data is not changed.

Debate de investigadores

0 investigadores lo mencionaron en los últimos 7 días

Publicaciones en X de investigadores de la lista de nday que mencionan este CVE, tal como se publicaron. nday no las verifica.

nday aún no ha recopilado publicaciones en X de investigadores de su lista sobre este CVE.

Discussion

0 comments

No comments yet.